ClearanceBrief
Recall Z-3127-2026 ยท OIB Blood And Urine Collection Kit (Excludes Hiv Testing)

Medline Industries, LP

Medline Convenience Kits: 1) DYNDB3086D, Catalog Number: DYNDB3086D; 2) DYNDB3100, Catalog Number: DYNDB3100; 3) DYNDH1619A, Catalog Number: DYNDH1619A; 4) DYNDH1620A, Catalog Number: DYNDH1620A; 5) DYNDH1694A, Catalog Number: DYNDH1694A; 6) DYNDH1695, Catalog Number: DYNDH1695; 7) DYNDH1708A, Catalog Number: DYNDH1708A; 8) DYNDH1714A, Catalog Number: DYNDH1714A; 9) DYNDH1723, Catalog Numb

DateSep 14, 2026 (initiated Aug 11, 2026)
StatusOpen, Classified
ReasonThe kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may lead to serious adverse health consequences, including systemic infection, sepsis, illness, and death.
Root causeNonconforming Material/Component
ActionAn URGENT MEDICAL DEVICE RECALL notification letter dated 8/11/26 was sent to customers. This communication serves as an update to the Medline notification sent on June 12, 2026. Please be advised that the actions required for certain item(s) and lot(s) identified in the original notification have changed based on additional recall details received from BD (Becton, Dickinson and Company). REQUIRED ACTION: 1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below
Quantity47848 kits
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.