Recall Z-3171-2026 ยท LLZ System, Image Processing, Radiological

GE Healthcare

Core Services Hub V8, a set of backend services including Cross Enterprise (XE) and Workflow Core (WFC) services that enables enterprise imaging workflows, All versions

DateSep 15, 2026 (initiated Jul 31, 2026)
StatusOpen, Classified
ReasonUnder certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.
Root causeUnder Investigation by firm
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/31/26 was sent to customers. Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue to use your systems by following the instructions below. 1. Do not perform any new inbound data (DICOM) migration projects. Please contact GE HealthCare before starting any new migration projects. 2. If using Enterprise Archive with Core Services Hub/Workflow Core Services, when reviewing or exporting prior imaging studies after migration to Enterprise Archive, verify that the study and images within the stud
Quantity250 systems
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043415CENTRICITY PACS SYSTEM
Ge Medical Systems Information Technologies, Jan 21, 2005
K110875CENTRICITY PACS
GE Healthcare, May 5, 2011