Recall Z-3169-2026 ยท LLZ System, Image Processing, Radiological
GE Healthcare
Centricity PACS Server V7, intended for managing digital medical images and associated data, All versions
| Date | Sep 15, 2026 (initiated Jul 31, 2026) |
| Status | Open, Classified |
| Reason | Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care. |
| Root cause | Under Investigation by firm |
| Action | An URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/31/26 was sent to customers. Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue to use your systems by following the instructions below. 1. Do not perform any new inbound data (DICOM) migration projects. Please contact GE HealthCare before starting any new migration projects. 2. If using Enterprise Archive with Core Services Hub/Workflow Core Services, when reviewing or exporting prior imaging studies after migration to Enterprise Archive, verify that the study and images within the stud |
| Quantity | 116 systems |
| Distribution | Worldwide - US Nationwide distribution. |
| FDA record | Open on fda.gov |