Recall Z-3334-2026 ยท HRS Plate, Fixation, Bone

Wright Medical Technology, Inc.

CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System

DateOct 1, 2026 (initiated Aug 18, 2026)
StatusOpen, Classified
ReasonScrews, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the plate's internal threads, which may lead to surgery prolongation to retrieve another device, or alternative surgery method if no replacement device is readily available.
Root causeProcess control
ActionOn 8/13/2026, recall notices were mailed and emailed to customers who were asked to do the following: 1) Review the product issue and risks as communicated in this Recall Notice and communicate the issue as appropriate in your organization. 2) Sign and return the enclosed business reply form by email to memphis.fieldaction@stryker.com 3) Upon receipt of the completed business reply form, firm will arrange for product return. 4) If you have further distributed the affected product, please notify the applicable parties at once about this recall. You may copy and distribute this notification l
Quantity34
DistributionUS Nationwide distribution in the states of FL, TX, PA, CA, ID, MI, OK, TN, NJ, WI, CO.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133715CHARLOTTE CLAW PLATE SYSTEM
Wrightmedicaltechnologyinc, Feb 28, 2014