Recall Z-3191-2026 ยท BSZ Gas-Machine, Anesthesia

Datex-Ohmeda, Inc.

Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000

DateSep 16, 2026 (initiated Jul 30, 2026)
StatusOpen, Classified
ReasonAnesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
Root causeNonconforming Material/Component
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/30/26 was sent to customers. Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue to use your Anesthesia system by following the instruction below: 1. Prior to each use, complete preoperative checkout and tests in accordance with the instructions in the User Reference Manual (URM). 2. When Alternate O2 control is enabled, the clinician should adjust the Alternate O2 flow as appropriate for the case. Use the Alternate O2 control knob to adjust the O2 flow. 3. When Alternate O2 control is en
Quantity7 units
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172045Aestiva 7900, Aestiva MRI, Aestiva 7100, Aestiva 7100 Compact, Aespire 100, Aespire 7100, Aespire 7900, Aespire View, Av
Datex-Ohmeda, Inc., Nov 3, 2017