Recall Z-3212-2026 · LHI Set, I.V. Fluid Transfer

ICU Medical, Inc.

CADD Yellow Medication Cassette Reservoir with NRFit connector US Product Codes: 21-7600-24, 21-7609-24 OUS Product Codes: 21-7609-24 21-7600-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24JP 21-7600-24JP 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7600-24 21-7600

DateSep 17, 2026 (initiated Aug 20, 2026)
StatusOpen, Classified
ReasonReservoirs may exhibit medication leakage at the joint between luer connector and tubing. Issue could result in delay or interruption of therapy, under delivery of medication, exposure to infectious or toxic agents, or air embolus.
Root causeProcess control
ActionOn August 20, 2026, the firm began notifying affected customers via Urgent Medical Device Correction letters. Customers were informed that certain reservoirs may exhibit fluid leakage at the joint between the connector and the tubing. Customers were instructed to check all inventory locations for affected product and discontinue use. All affected products should be destroyed following their institution's destruction process. If destruction is not immediately possible at your facility, quarantine the products until disposal. Customers should share the notification with all potential users of
Quantity185,688 units
DistributionWorldwide - US Nationwide distribution in the states of AR, CA, FL, IL, IN, LA, MI, MO, MS, TN, TX, VA, WI and the countries of Austria, Australia, Belgium, Costa Rica, Germany, Estonia, Spain, Finland, France, Great Britain, Ireland, Italy, Japan, Kuwait, Netherlands, Norway, and Sweden.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162219CADD® Infusion Disposables Portfolio with NRFit™ connectors
Smiths Medical Asd, Inc., Jul 20, 2017