Recall Z-3143-2026 ยท DQK Computer, Diagnostic, Programmable

GE Medical Systems Information Technologies Inc

Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV

DateSep 11, 2026 (initiated Jul 30, 2026)
StatusOpen, Classified
ReasonDevice may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.
Root causeUnder Investigation by firm
ActionOn July 30, 2026, the firm sent Urgent Medical Device Correction letters to customers with affected units. This letter informs the customer of the issue and provides actions to be taken by the customer/user for continued use of the device. GE HealthCare will provide a User Manual Addendum that provides guidance regarding the prolonged startup condition.
Quantity29,570 (23,632 US; 5,938 OUS)
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162012CARESCAPE Central Station V2
GE Healthcare, Sep 15, 2016