ClearanceBrief
Recall Z-3185-2026 ยท NLM Laparoscope, General & Plastic Surgery, Reprocessed

Stryker Sustainability Solutions

REF:B12LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 12X100MM

DateSep 16, 2026 (initiated Aug 19, 2026)
StatusOpen, Classified
ReasonReusable access port system may crack or seperate during use.
Root causeDevice Design
ActionOn August 19, 2026, Stryker Sustainability Solutions (SSS)issued a Urgent: Medical Device recall notice to affected consignees via sales representative and US Mail. SSS ask consignees to take the following actions: 1. Inform individuals within your organization who need to be aware of this action. 2. Immediately check your internal inventory to locate the affected product. Segregate the affected products, quarantine and discontinue use. 3. Complete the Business Reply Form or use the QR Code on Attachment A. Response is required. a.If you have indicated that product remains in your inventory,
Quantity2682 units
DistributionUS Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Mis
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201511Reprocessed Endoscopic Trocars and Sleeves
Stryker Sustainability Solutions, Jul 31, 2020