Recall Z-3202-2026 ยท LGE Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Trunnion-Bearing

Howmedica Osteonics Corp.

BEARING AND AXLE BUSHING ASSY STD ELBOW C-M113-1-301 KRH REV TIB INS 16mm THK .640 DIA STEM C-M113-5-600 SELF-PRESSURIZING GLENOID SIZE 9 C-M063-5-800 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

DateSep 16, 2026 (initiated Jul 30, 2026)
StatusOpen, Classified
ReasonAffected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Root causeProcess change control
ActionFirm began notifying customers on July 30, 2026, via FedEx 2-day air. Customers were instructed to quarantine and discontinue use of the affected product. Affected product should be returned to the firm.
Quantity12,156 US
DistributionWorldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.