Recall Z-3201-2026 · KWZ Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer

Howmedica Osteonics Corp.

ACETABULAR CUP UHMWPE 61-3258 ALL POLY CONSTRAINED INSERT 44MM 69-2244 ALL POLY CONSTRAINED INSERT 46MM 69-2246 ALL POLY CONSTRAINED INSERT 48MM 69-2248 ALL POLY CONSTRAINED INSERT 50MM 69-2850 ALL POLY CONSTRAINED INSERT 52MM 69-2852 CONSTRAINED ACETABULAR INSERTS 22MM 50MM 2099-2250 CONSTRAINED ACETABULAR INSERTS 22MM 52MM 2099-2252 CONSTRAINED ACETABULAR INSERTS

DateSep 16, 2026 (initiated Jul 30, 2026)
StatusOpen, Classified
ReasonAffected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Root causeProcess change control
ActionFirm began notifying customers on July 30, 2026, via FedEx 2-day air. Customers were instructed to quarantine and discontinue use of the affected product. Affected product should be returned to the firm.
Quantity12,156 US
DistributionWorldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K261624Omnifit® 20° Series I Insert;Omnifit® Crossfire® 10° Series II Insert; Constrained Acetabular Insert; Trident® Crossfire
Howmedica Osteonics Corp., Dba Stryker Orthopaedics, Jul 10, 2026