ClearanceBrief
Recall Z-3246-2026 ยท KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories)

Olympus Corporation of the Americas

Olympus BiCOAG Hemostasis Probe 7Fr Coaxial Connector. Model/Catalog Number: CD-B620LA.

DateSep 18, 2026 (initiated Aug 19, 2026)
StatusOpen, Classified
ReasonPotential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.
Root causeNonconforming Material/Component
ActionOlumpus notified consignees on 08/19/2026 via "URGENT: MEDICAL DEVICE REMOVAL" letter. Consignees were informed of the reason for recall and hazard involved. Consignees were instructed to examine inventory for affected units, immediately cease use and quarantine them, arrange for their return, and confirm receipt of the recall notification through the portal. Consignees were also instructed to notify customers if product has been further distributed.
Quantity2,142 units
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Latin America, Singapore, Hong Kong, and EU.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123319GYRUS ACMI BICOAG HEMOSTASIS PROBE
Olympus Surgical Technologies America, Dec 14, 2012