Recall Z-3262-2026 ยท IOQ Bed, Flotation Therapy, Powered
Baxter Healthcare Corporation
PRO+ MRS SURFACE FOR VERSACARE, mattress with Advanced Microclimate Technology (AMT) uses a non-powered system of intake and output valves to provide weight-based pressure redistribution whenever a patient moves or is repositioned, Model/Catalog Number: P3255A08
| Date | Sep 24, 2026 (initiated Aug 18, 2026) |
| Status | Open, Classified |
| Reason | There is an error in the product description on the Unique Device Identification (UDI) label. These discrepancies in labeling do not align with the approved product or product specifications and may lead to incorrect product identification or selection, potentially resulting in use with an incompatible bed frame. |
| Root cause | Process control |
| Action | An "Urgent Medical Device Correction notification" letter dated 8/18/26 was sent to customers. Actions to be Taken by Customers 1.Locate the affected product at your facility and confirm each Product Code. Once product has been identified, use Hillrom pro+ Surface Instructions for Use (209196) (IFU) to assess the fit of the surfaces relative to the frame and siderails to confirm the surface is installed properly. If the surface does not meet the requirements set in the IFU please discontinue the use of this surface. 2. A Baxter representative will contact your facility to replace the incorrec |
| Quantity | 169 units |
| Distribution | Worldwide distribution - US Nationwide and the country of Canada. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.