Recall Z-3033-2026 ยท FQO Table, Operating-Room, Ac-Powered

IMRIS Imaging Inc

ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000

DateAug 26, 2026 (initiated Aug 6, 2026)
StatusOpen, Classified
ReasonThe linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.
Root causeNonconforming Material/Component
ActionAn URGENT: MEDICAL DEVICE RECALL notification letter dated 8/6/26 was sent to customers. IMRIS recommends the following preventative actions until the corrective actions have been implemented on your table: " Per the OR Table Operator Manuals, the table should be tested daily in all standard positions and inspected for anomalies including the fit of the covers, tabletops, headrest, pendant connection and rotation. " During usage, verify the table position is locked by manually pushing the table to ensure it does not exhibit rotational movement. " If the floor latching mechanism is observed to
Quantity6 units
DistributionWorldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden.
FDA recordOpen on fda.gov