Recall Z-3333-2026 ยท EOB Nasopharyngoscope (Flexible Or Rigid)

Karl Storz Endoscopy

HOPKINS Telescope 30, 4mm REF 10005BA UDI-DI code: 04048551045121 KARL STORZ HOPKINS and HOPKINS II telescopes are to be used for visualization of the surgical site during diagnostic and therapeutic endoscopic procedures.

DateOct 1, 2026 (initiated Aug 12, 2026)
StatusOpen, Classified
ReasonDue to incorrect inside packaging label for eyepiece.
Root causeEmployee error
ActionOn 08/12/2026, an "URGENT: MEDICAL DEVICE RECALL" Letter was sent via email informing customers that, KARL STORZ distributed 6 HOPKINS Telescope 30, 4mm REF: 10005BA between November 24, 2025, and December 15, 2025, that contains an optic (eyepiece) that is incorrectly etched with REF: 10005AA, which is a HOPKINS Telescope 0, 4 mm. The optic (eyepiece) inside the packaging is in fact a REF: 10005BA HOPKINS Telescope 30, 4mm, as the outer packaging indicates. Customers are instructed to: 1. Immediately check for impacted serial numbers in inventory and quarantine the incorrectly etched devices
Quantity6 eyepieces
DistributionU.S. Nationwide distribution in the states of AZ, CA and MO.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K945788KARL STORZ RIGID TELESCOPES FOR ENT ENDOSCOPIC PROCEDURES
KARL STORZ Endoscopy-America, Inc., Jan 25, 1995