Recall Z-3070-2026 ยท LRO General Surgery Tray
Medline Industries, LP
Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators BLOOD CULTURE PINK PERIPHERAL DYNDV1875 BOTTOM DRAWER ACC010972 DRAWER #4 ACC010479A DRAWER 2 ACC010690 ACC010875 DRAWER 3 SYR/DRS/MON/NG 1 OF 3 ACC011073 DRAWER 4 CIRCULATION ACC011088 DRAWER 5 - LEFT BOX ACC011067 LABOR TRIAGE KIT - LF #13232 DYKM1564 DYKM1564A LINEN UNIVERSAL PA
| Date | Aug 31, 2026 (initiated Jul 20, 2026) |
| Status | Open, Classified |
| Reason | Downstream recall - manufacturer recalled specific lots of BD ChloraPrep Clear - 1mL Applicators. Affected product may exhibit an open or incomplete seal on the packaging of the applicator. Use of potentially contaminated applicator may lead to localized skin or soft tissue infection, or systemic infection or infectious complications. |
| Root cause | Nonconforming Material/Component |
| Action | Firm began notifying consignees on 7/20/2026 via First-Class Mail and email. Medline Industries LP instructed customers to request stickers to over-label the affected convenience kits with a label indicating to remove the affected component and replace it with product from supply. No kits are to be returned. |
| Quantity | 222,653 total |
| Distribution | US Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K213481 | Medline General Surgery Tray Medline Industries, Inc., May 24, 2022 |