ClearanceBrief
Recall Z-3159-2026 ยท ONB Closed Antineoplastic And Hazardous Drug Reconstitution And Transfer System

ICU Medical, Inc.

Oncology sets with drip chambers

DateSep 11, 2026 (initiated May 12, 2026)
StatusOpen, Classified
ReasonIntravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.
Root causeProcess control
ActionOn 5/12/2026, the firm posted a Information Regarding Reported Black Specks in Certain IV Tubing Set Drip Chambers from ICU Medical document on their website. On 5/18/2026, updated the notices were sent. On 7/28/2026, correction notices were sent to customers asking them to do the following: 1) Share this notification with all potential users of the device to ensure they are aware of this notification. If the devices are used or further distributed to another location, please ensure that this communication is delivered there. 2) Complete and return the attached Customer Response Form to M
Quantity28,550
DistributionWorldwide - US Nationwide distribution including in the states and territories of ND, TX, CT, KY, CO, AR, NC, NJ, VA, MD, OK, AK, IA, IN, KS, OR, CA, OH, WA, GA, AL, SC, MA, HI, FL, NH, NE, MN, MS, SD, MT, NY, DC, ID, WI, PA, AZ, MO, NV, TN, NM, WV, WY, IL, UT, MI, RI, VT, LA, DE, ME, VI, GU, PR and
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131549CHEMOLOCK, CHEMOLOCK PORT
Icu Medical, Inc., Jul 26, 2013
K173477ChemoCLAVE Cytotoxic Medication Preparation and Delivery System
Icu Medical, Inc., Aug 31, 2018