Recall Z-3066-2026 ยท FRN Pump, Infusion

Baxter Healthcare Corporation

Baxter Novum IQ Syringe Pump, Loaner, Model Number 40800BAXUSL

DateAug 31, 2026 (initiated Aug 11, 2026)
StatusOpen, Classified
ReasonA software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be Infused (VTBI) remains.
Root causeSoftware design
ActionOn August 11, 2026, firm began sending customers Urgent Medical Device Correction delivered via email and USPS first class mail. Customers were advised to use the same compatible syringe model (brand/size) for all syringes required within a Multi-Syringe VOT workflow. Customers were provided with additional instructions for situations where the syringe does not support the intended flow rate and the user receives a prompt to confirm continuation of the infusion at the maximum or minimum supported flow rate for that syringe after a syringe change.
Quantity4 units
DistributionUS Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K242390Novum IQ Syringe Pump (40800BAXUS)
Baxter Healthcare Corporation, Sep 5, 2024