ClearanceBrief
Recall Z-3097-2026 ยท NWU Endoscope Introducer Kit

Cardinal Health 200, LLC

Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT,ROBOTICS DH , Catalog Number: PGOCRDNHD; 2) KIT, EAS ROBOTIC , Catalog Number: PG36RPEAB; 3) KIT, VIDEO THORACOSCOPY CMC , Catalog Number: PVCGVTACF

DateSep 3, 2026 (initiated Jul 29, 2026)
StatusOpen, Classified
ReasonThe Urinary Catheter Trays contain CliniClean Pericare Wipes which may potentially be contaminated with a lubricant used during the manufacturing process. If a contaminated wipe is used on a patient, localized skin effects may occur. These may include irritation, erythema, or mild inflammation.
Root causeProcess change control
ActionAn URGENT: PRODUCT CORRECTION notification letter dated 7/29/26 was sent to customers. If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits. 2. ALERT your clinicians of this action. 3. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard the recalled component(s). 4. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Additional labels will be sent to you
Quantity115 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.