Recall Z-2984-2026 ยท MPB Catheter, Hemodialysis, Non-Implanted
Bard Access Systems, Inc.
Power-Trialysis Short-Term Straight Dialysis Catheter, REF: 5605150; Power-Trialysis Short-Term Straight Dialysis Catheter, REF: 5605200; Power-Trialysis Short-Term Straight Dialysis Catheter, REF: 5605240; Power-Trialysis Short-Term Straight Dialysis Catheter, REF: 5605300; Power-Trialysis Short-Term Curved Extension Dialysis Catheter, REF: 5615120; Power-Trialysis Short-Term Curved Extension D
| Date | Sep 11, 2026 (initiated Aug 10, 2026) |
| Status | Open, Classified |
| Reason | this is a downstream recall of class I recall RES 98777 |
| Root cause | Material/Component Contamination |
| Action | On 8/10/2026, recall notices were mailed or emailed to customers and distributors who were asked to do the following: 1) Do not use the Sodium chloride ampules contained within the affected products. The sodium chloride ampule is the only impacted component within the tray and all remaining components can continue to be used. 2) Secure an alternative local Sodium chloride prior to entering the procedural environment. 3) Upon opening the tray, remove and destroy the ampules within a controlled and sterile environment to maintain the sterility of the remaining components within the tray. 4) |
| Quantity | 149521 |
| Distribution | US: FL, IL, OH, VA, MD, TX, WI, GA, TN, IN, SC, NC, MI, PA, OR, OK, AZ, MS, NJ, AL, CA, CO, MA, NV, ID, MO, LA, WA, KY, IA, NH, NY, KS, HI, NE, MN, MT, WY, AR, RI, WV, UT, PR, DE, CT, ME, DC, ND, NM, SD |
| FDA record | Open on fda.gov |