Recall Z-3047-2026 ยท GGP Test, Qualitative And Quantitative Factor Deficiency

Diagnostica Stago, Inc.

STA Staclot Protein S, Model/Catalog Number: 00746-Reagent Lot number 273074; qualitative and quantitative factor deficiency test

DateAug 28, 2026 (initiated Jul 29, 2026)
StatusOpen, Classified
ReasonA positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified
Root causeUnder Investigation by firm
ActionDiagnostica Stago issued an URGENT: MEDICAL DEVICE RECALL (REMOVAL) notice to its consignees on 7/29/2026 via email. The notice described the issue, risk assessment, and requested the following: "Actions Required To prevent potential workflow disruptions associated with the increased recalibration frequency, we request that you: " Immediately discontinue use of STA-Staclot Protein S, Lot 273074. " Dispose of any remaining inventory of the affected lot in accordance with your facility's procedures and local regulations. " Complete, sign, and return the enclosed Acknowledgement Form by email
Quantity5220 units
DistributionUS Nationwide distribution in the states and territory of NJ, NY, NJ, VA, PA, FL, AL, GA, TN, OH, IL, MO, NE, TX, CO, UT, CA, PR.