Recall Z-3030-2026 ยท LLZ System, Image Processing, Radiological

FUJIFILM Healthcare Americas Corporation

FUJIFILM Synapse PACS software v7.4.310; radiological image processing system

DateAug 26, 2026 (initiated Jul 22, 2026)
StatusOpen, Classified
ReasonSoftware issue may cause the ruler measurement to be inaccurate.
Root causeSoftware design
ActionFUJIFILM issued an URGENT MEDICAL DEVICE CORRECTION to its consignees on 7/21/2026 via email. The notice explained the issue and how to identify it, risk to health, and requested the following: "ACTIONS TO BE TAKEN BY THE CUSTOMER/END-USER 1. Please post this notification on or near the Synapse PACs workstation. 2. Please complete and return the Field Action Verification Form attached to this communication. Responding with the information requested is essential for ensuring appropriate action is taken. 3. Please contact Technical Assistance Center (TAC) at 1-888-FUJIMED (1-888-385
Quantity6 units
DistributionWorldwide - US Nationwide distribution in the states of NY, VA, KS, OR, OH, and the country of Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190232Synapse PACS
Fujifilm Corporation, Mar 26, 2019