ClearanceBrief
Recall Z-3083-2026 ยท MXD Recorder, Event, Implantable Cardiac, (With Arrhythmia Detection)

Boston Scientific Corporation

LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS ANZ, Model Number 7248, All Versions (Current 7.7)

DateSep 2, 2026 (initiated Aug 6, 2026)
StatusOpen, Classified
ReasonThere is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs). Delayed notification of an AF Burden Alert may delay clinical evaluation or therapeutic intervention of an atrial arrhythmia.
Root causeSoftware design
ActionAn Urgent Medical Device Correction dated August 2026 was sent to customers. Boston Scientific is notifying you of the potential for a delay in AF Burden Alert notification affecting the LATITUDE" Clarity server software which supports the LUX-Dx" family of ICMs. This delayed AF Burden Alert notification behavior by LATITUDE Clarity may occur if the Alert for AF Burden e Threshold is configured to ">0 hours per day" setting (a non-default configuration). The ICM is performing as intended with continued AF detection, data storage, and transmission of device data. Boston Scientific has received
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K231328LUX-Dx II (M302); LUX-Dx II+ (M312)
Boston Scientific Corp, Aug 19, 2023