Recall Z-3178-2026 ยท OYK Ankle Arthroplasty Implantation System

restor3d Inc.

Kinos Axiom Total Ankle System with Axiom PSR, Flat-cut (Cut Guide), REF: 5831-9006

DateSep 16, 2026 (initiated Jul 27, 2026)
StatusOpen, Classified
ReasonAnkle implantation systems were packaged with incorrect patient-specific cutting guides, which may result in: delayed surgery, the need to use standard instrumentation, reduced fixation or poor alignment of implants, implant loosening/migration, or need for revision surgery.
Root causeLabeling Change Control
ActionOn 7/28/2026, recall notices were emailed to customers who were asked to do the following: 1) Segregate and return affected devices. 2) Complete and return the response form. If additional questions, contact firm at 984-201-2310
Quantity2
DistributionUS Nationwide distribution in the states of FL, CO.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K223326Axiom PSR System
Restor3D, Mar 30, 2023