Recall Z-2999-2026 ยท OFR Tracheobronchial Suction Catheter Kit

Medline Industries, LP

Medline medical convenience kit labeled as: TRACH TUBE KIT, Medline Kit SKU DYKM2330

DateAug 21, 2026 (initiated Jul 10, 2026)
StatusOpen, Classified
ReasonComponent labeling discrepancy
Root causeLabeling Change Control
ActionMedline Industries issued an URGENT MEDICAL DEVICE CORRECTION notice to its sole consignee on 7/10/2026 via USPS and email. The notice explained the issue, potential risk, and attached the notice from Medtronic. The notice from Medtronic provided the following information: "Customer Actions: " Customers can continue to use the Covidien Shiley" disposable decannulation plugs. " Post this notice in all care environments where the affected products are used to maintain awareness to this labeling issue. " Share this notice with all relevant personnel within your organization, and with any orga
Quantity45 kits
DistributionUS Nationwide distribution in the state of California.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.