ClearanceBrief
Recall Z-2994-2026 ยท OGR Ear, Nose, And Throat Surgical Tray

Medline Industries, LP

Medline medical convenience kits, labeled as: 1) SINUS PACK-LF, Medline Kit SKU DYNJ0966146F; 2) ENT PACK, Medline Kit SKU DYNJ40154D; 3) ENT PACK, Medline Kit SKU DYNJ40154F; 4) ENT SCOPE PACK, Medline Kit SKU DYNJ44480D; 5) ENT NASAL ORAL PACK MOSES, Medline Kit SKU DYNJ58928A; 6) HEAD & NECK PACK, Medline Kit SKU DYNJ83296; 7) TRANSPHENOIDAL PACK, Medline Kit SKU DYNJ89006; 8) SINUS, M

DateAug 21, 2026 (initiated Jul 8, 2026)
StatusOpen, Classified
ReasonSpecific item(s) and lot(s) of Medline convenience kits containing Pilling Wecksorb Neurosponges and Neurosponge Strips are being recalled due to potential out-of-specification levels of endotoxin of the neurosponges.
Root causeNonconforming Material/Component
ActionMedline Industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 7/8/2026 via USPS and email. The notice explained the issue, potential adverse health consequences, and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected p
Quantity2262 kits
DistributionUS Nationwide distribution in the states of AZ, CA, IA, ID, IL, IN, MA, MD, MI, MN, MO, MT, NY, OH, TX, WI, WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.