Recall Z-2872-2026 ยท FMI Needle, Hypodermic, Single Lumen

Bard Access Systems, Inc.

BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK

DateAug 14, 2026 (initiated Jul 30, 2026)
StatusOpen, Classified
ReasonAfter placement of intraosseous (IO) needle sets, used when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases, users may experience difficulty/inability to remove obturator, due to it locking to the needle hub, due to out-of-tolerance dimensions and IFU unclarity, which may result in new needle placement, which may result in delayed therapy
Root causeComponent change control
ActionOn 7/30/2026, recall notices were mailed/emailed: If rotation and locking occurs, the clinician may need to obtain a new IO needle or other form of vascular access device, such as a Peripheral IV catheter. Alternate vascular access approaches, including additional peripheral IV attempts, or central venous access may continue to be pursued based on clinical judgment/protocols. Affected product should not be used to establish IO access. For patients in whom potentially affected intraosseous needles are currently in use/or have been used and are/were functioning properly, no additional steps
Quantity80,200
DistributionUS: MA, FL, IA, SC, IL, NY, UT, NJ, ID, IN, PA, CA, TX, OH, MO, VA, MD, NC, AL, MN, CO, KY, LA, NV, OK, MS, AZ, MI, GA, AR, KS, TN. OUS: CA, SG
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191976Piper GO-IO Intraosseous Infusion System
Piper Access, LLC, Nov 13, 2019
K203193BD Intraosseous Infusion System
Bard Access Systems, Inc(Bas)[Wholly-Owned Subsidiary of Bd], Mar 4, 2021