Recall Z-3045-2026 ยท OJG Neurological Tray
AVID Medical, Inc.
1) NEURO, (kit) ACNE73-01; 2) NEURO, (kit) ACNE73-03; 3) CRANI, (kit) AKCR19-01; 4) ORTHO SPINE PACK, (kit) ARMC017; 5) BMHM NEURO PACK, (kit) BMHT016-05; 6) LUMBAR PACK, (kit) BMHT112-03; 7) NEURO LAMINECTOMY PACK, (kit) DGNE75; 8) NEURO LAMINECTOMY PACK, (kit) DGNE75-02; 9) DRAH NEURO LAMINECTOMY SNEDHDN,(kit) DRNE75; 10) DRAH NEURO LAMINECTOMY SNEDHDN, (kit) DRNE75-01; 11) CRANIOSYNOS
| Date | Aug 28, 2026 (initiated May 19, 2026) |
| Status | Open, Classified |
| Reason | The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected variability of endotoxin levels on the products. |
| Root cause | Material/Component Contamination |
| Action | Owens & Minor issued an URGENT: MEDICAL DEVICE RECALL NOTIFICATION to its consignees on 5/19/2026 via email. The notice explained the issue, potential risk to health and requested the following: "ACTIONS TO TAKE: " Identify, segregate, and quarantine all impacted product. " Add recall labels to affected Avid kits to ensure all users are aware of the actions to discard the impacted component. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. Please contact your Sales Representative if pre-printed labels are needed. This |
| Quantity | 16223 kits |
| Distribution | US Nationwide distribution in the states of AZ, CA, CO, WA, WY, TX, MI, FL, MD, UT, NC, MS, PA, MO, IA, NE, OH, IL, SD, MN, GA. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.