Biofire Defense

FilmArray NGDS Warrior Panel REF NGDS-ASY-0007 UDI-DI code: (01)00851458005136 The FilmArray NGDS Warrior Panel is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with the FilmArray 2.0 system. The FilmArray NGDS Warrior Panel detects and identifies Bacillus anthracis, Yersinia pestis, Francisella tularensis, Coxiella burnetii, Ebola virus, and Marburg vir

DateAug 4, 2026 (initiated Jul 16, 2026)
StatusOpen, Classified
ReasonFirm has identified mismatches between the outbreak strain and the reference strain used in the development of its qualitative, multiplexed, nucleic acid-based in vitro diagnostic panel test.
Root causeUnder Investigation by firm
ActionOn 07/16/2026, an "ADVISORY NOTICE URGENT: MEDICAL DEVICE CORRECTION" Letter was sent via FedEx Priority Overnight to customer informing them that, due to the emergence of the 2026 Bundibugyo outbreak strain, BioFire Defense has conducted testing to evaluate performance of the Orthoebolavirus spp. assays on the FilmArray NGDS Warrior Panel. This preliminary testing has identified a potential reduction in sensitivity to the 2026 Bundibugyo outbreak strain which could result in Orthoebolavirus spp. Not Detected results when testing on the FilmArray NGDS Warrior Panel. Customers are instructed
Quantity122 kits (3,660 pouches)
DistributionU.S. Nationwide distribution in the states of AL, CA, DC, FL, GA, HI, KY, LA, MA, MD, MN, NC, NH, NV, NY, OH, RI, TN, TX, UT, VA and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170883FilmArray NGDS Warrior Panel
Biofire Defense, LLC, Jun 22, 2017