Recall Z-2927-2026 ยท GTJ Antisera, All Groups, N. Meningitidis

Remel Europe Ltd.

Thermo Fisher Diagnostics Neisseria meningitidis Gp C 2ml/VL, Product Number R30166901; invitro diagnostics

DateAug 7, 2026 (initiated Jul 17, 2026)
StatusOpen, Classified
ReasonCertain lots of Neisseria meningitidis Gp C 2ml/VL, R30166901, are not performing to IFU criteria not positive to ATCC13102 within 60 Seconds.
Root causeUnder Investigation by firm
ActionThermo Fisher Scientific issued an Urgent: Medical Device Recall notice to its consignees on 7/17/2026 via USPS, first class mail. The notice explained the issue, risk to health, and requested the following: "ACTIONS TO BE TAKEN BY THE CUSTOMER: Our records indicate that you have received above products. Accordingly, in keeping with our Quality Policy, we request that you please notify any personnel who need to be aware of this issue. Laboratory professionals are advised to destroy any remaining inventory of lots listed above. Please complete the accompanying Acknowledgement form and return
Quantity26 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.