Recall Z-2964-2026 ยท OKV Obstetrical Kit

AVID Medical, Inc.

Halyard LAVH PACK, Model/Catalog Number: USAR053-11

DateAug 14, 2026 (initiated Jul 10, 2026)
StatusOpen, Classified
ReasonA customer complaint reported that the Clearify Visualization System (5303907) leaked in a kit. As a result of Avid s investigation, it was determined that the component was subjected to EO sterilization which is not in alignment with the manufacturer s recommendation.
Root causeProcess control
ActionOwens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION notice to its consignees on 7/10/2026 via email. The notice explained the issue, potential risks to health, and requested the following: "ACTIONS TO TAKE: " DO NOT USE AND DISCARD the Clearify Visualization System (Component# 5303907) located in the convenience kit identified in Table 1 above. All other components in the kits may continue to be used. " Add warning labels to affected AVID kits to ensure all users are aware of the actions to take prior to use. A labeling template is enclosed for printing labels which can be a
Quantity40 packs
DistributionUS Nationwide distribution in the state of TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.