Recall Z-2933-2026 ยท HIB Speculum, Vaginal, Nonmetal

OBP Medical Corporation

OBP Medical , ER-Spec Single-Use Vaginal Speculum with LED Light Source, X-Small REF: C020130-1

DateAug 10, 2026 (initiated Jul 13, 2026)
StatusOpen, Classified
ReasonThere is a potential for vaginal speculums to contain a burr on the proximal plastic blade, which may introduce a sharp edge in the contact regions for both the patient and the healthcare provider.
Root causeProcess control
ActionOn July 14, 2026, OBP Medical a subsidiary of Cooper Surgical issued a Urgent Medical Device Recall Notification to affected consignees via UPS. OBP Medical asked consignees to take the following actions: 1. Physicians/Facilities: Please remove all product of the subject lots from your inventory and note the number of units to be returned on the attached Recall Response Form. 2. Distributors: Please remove all product of the subject lots from your inventory and note the number of units to be returned on the attached Distributor Only Acknowledgement Form. Per the instructions, please commun
Quantity8,224 boxes (148,032 units)
DistributionWorldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV and the countries of HKG, CAN, NOR, UAE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K091937OFFICE-SPEC DISPOSABLE SIDE-OPENING SPECULUM
Obp Corporation, Sep 9, 2009