Recall Z-2959-2026 ยท ILQ System, Communication, Powered

Tobii Dynavox Llc

TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798

DateAug 14, 2026 (initiated Jul 8, 2026)
StatusOpen, Classified
ReasonUnder certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
Root causeDevice Design
ActionOn July 8, 2026, the firm notified customers by emailing letters titled "URGENT: MEDICAL DEVICE RECALL NOTICE." Customers were informed that Tobii Dynavox would provide a repair authorization number (RA#), prepaid shipping label, inspection, repair, and return shipping for the affected device. The repair will include modifications intended to improve battery stability within the device. Customers were provided with ways to contact the company to obtain a repair authorization number. Customers were informed that it is safe to continue using the device as intended while waiting for the repair
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.