Recall Z-2921-2026 ยท GAW Suture, Nonabsorbable, Synthetic, Polypropylene

Covidien, LP

Surgipro SP-623 SURGIPRO* 2-0 BLU 75CM SC2 X36 SP-623

DateAug 7, 2026 (initiated Jun 24, 2026)
StatusOpen, Classified
ReasonProduct may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or prolonged surgery.
Root causeEquipment maintenance
ActionThe firm began notifying consignees on 7/7/26 via email, mail, or electronically. Notices were titled, "URGENT: VOLUNTARY MEDICAL DEVICE RECALL." Customers were informed of the manufacturing nonconformance. If products in scope of this field action have been used in patients under the customer's care, the customer is instructed to monitor for signs of infection or vision issues due to the possibility of biocontamination. Customers were instructed to immediately identify, cease use, and quarantine all unused impacted product. Unused impacted product is to be returned to Medtronic. The reca
Quantity135,875 total
DistributionUS Nationwide and the countries of Australia, Austria, Belgium, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.