Recall Z-2934-2026 · KRO Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer

Waldemar Link GmbH & Co. KG (Mfg Site)

Endo-Model, Replacement Set, Tibial Plateau, right/left, L, UHMWPE, CoCrMo, Rotating Hinge Item Number 15-0027/16

DateAug 11, 2026 (initiated Jul 6, 2026)
StatusOpen, Classified
ReasonThe affected Endo-Model replacement set for a Rotating Hinge tibial plateau incorrectly contains a Pure Hinge plateau. The incorrect plateau could lead to a prolonged or modified surgery as the plateau cannot be inserted and an alternative product would be needed. If the incorrect plateau were to be implanted, this is associated with increased risk of wear, which could result in a reduced standing time of the implant.
Root causeProcess change control
ActionForeign manufacturer, Waldemar Link GmbH & Co KG, notified importer LinkBio Corp. on 7/6/2026 via email. LinkBio then notified the single US consignee by email/Docusign on 7/7/2026. Customer was instructed to return the affected product to LinkBio.
Quantity1 US; 9 OUS
DistributionDomestic US distribution to Indiana. OUS distribution to Belarus and Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143179The LINK® Endo-Model® Knee System
Waldemar Link GmbH & Co. KG, Jan 30, 2015