Recall Z-2958-2026 ยท HTY Pin, Fixation, Smooth

Arrowhead DE LLC

Arrow-LOK Implant, REF: 21-253519, 21-253522, 21-253525, 21-253528, 21-253530, 21-253533, 21-253535, 21-253540, 21-253545, 21-253550, 21-353513, 21-353516, 27-253519, 27-253522, 27-253525, 27-253528, 27-253530, 27-253533, 27-253535.

DateAug 14, 2026 (initiated Jun 23, 2026)
StatusOpen, Classified
ReasonImplant indicated for fixation of osteotomies, arthrodeses and reconstruction in the lesser toes following corrective procedures, did not undergo required gamma radiation dose audit. A potential reduction in sterility assurance could lead to patient infection.
Root causeProcess change control
ActionOn 6/23/2026 customers were phoned and informed to return affected implants. On 7/7/2026, recall notices were mailed to risk managers who were asked to do the following: 1) Cease use, distribution, and quarantine affected product. 2) Contact firm customer service to obtain instructions for returning product. 3) Email mail tracking numbers to firm. 4) Complete and return the acknowledgement and receipt form. Questions can be directed to the firm at 470-419-1429
Quantity4,171
DistributionUS Nationwide distribution in the states of VT, FL, TN, NY, GA, TX, HI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112675ARROW-LOK DIGITAL SYSTEM
Arrowhead Medical Device Technologies, LLC, Dec 12, 2011