Recall Z-2745-2026 ยท RFN Fiber Optic Communication And Data Transfer

Terex Utilities

Power Over Fiber Base Module

DateJul 30, 2026 (initiated May 11, 2026)
StatusOpen, Classified
ReasonDevice not certified - no laser product report was filed with CDRH prior to distribution. Product label incorrectly states device is "Class IV", although Class 1 is appropriate for device. Improperly labeled - omitted labels required under 21 CFR 1040.10 / IEC 60825-1 via LN#56. Missing certification and proper manufacturer's ID label. Was not accompanied by compliant User Information. Formalized quality control procedures and record keeping of QC testing was not performed to IEC 60825-1 nor 21 CFR 1040.10 standards. Sales information did not convey required class designation and warning st
Root causeRadiation Control for Health and Safety Act
ActionA letter was received stating the product, problem and actions to be taken. The following actions being taken include: A. Develop and carry out a process to remove incorrect labeling and install all labels required under IEC 60825-1 2014. B. Generate, detail & submit full CAP (Corrective Action Plan) for FDA approval C. Generate initial laser product report or supplemental as appropriate. D. Equipment update to be performed by Terex personnel at customer location E. Procedures: * Fully certify device as Class 1 under the guidelines of FDA Laser Notice #56 by applying the IEC 60825-1 2014 lase
Quantity101
DistributionU.S. Nationwide Distribution and International Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.