Recall Z-2881-2026 ยท LFG Radioimmunoassay, Tricyclic Antidepressant Drugs

Siemens Healthcare Diagnostics, Inc.

Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL

DateAug 4, 2026 (initiated Jun 26, 2026)
StatusOpen, Classified
Reasonout-of-specification sodium azide concentration in Syva Emit Tox Serum Tricyclic Antidepressant Calibrator/Controls Lot U2. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.
Root causeUnder Investigation by firm
ActionOn June 26, 2026, the firm began notifying customers via Urgent Medical Device Correction letters. Customers were informed of the issue and provided with revised Safety Data Sheets (SDS) specific to the impacted lots for handling and disposal information.
Quantity494 US; 114 OUS
DistributionWorldwide - US Nationwide distribution in the states CA, ID, IL, IN, MA, MS, NC, NM, NY, OH, OK, OR, PA, WA, and Northern Mariana Islands part of USA and the countries of Canada, South Africa, Brazil, Chile, Colombia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K821743EMIT-TOX SERUM TRICYCLIC ANTIDEPR. ASSAY
Syva Co., Jul 2, 1982