Recall Z-2980-2026 ยท QFG Alternate Controller Enabled Insulin Infusion Pump

Medtronic MiniMed, Inc.

MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712;

DateAug 18, 2026 (initiated Jul 13, 2026)
StatusOpen, Classified
ReasonLoaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
Root causeProcess design
ActionOn 7/13/2026, field safety notices were sent to customers asking them to do the following: 1) Stop using the loaner pump and contact the firm to arrange for loaner pump return. 2) Contact your Healthcare Professional (HCP) to ensure you have access to backup insulin therapy supplies or alternative plans, which may include short- and long-acting insulins, syringes, with dosage instructions from your HCP. 3) Complete and return the customer acknowledgement form. Questions can be directed to the firm's Tech Support: 1-800-646-4633
Quantity16
DistributionInternational distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K251032MiniMed 780G insulin pump
Medtronic Minimed, Inc., Jul 1, 2025