Recall Z-2983-2026 ยท FLE Sterilizer, Steam

Fort Defiance Industries, LLC

Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

DateAug 18, 2026 (initiated Jul 8, 2026)
StatusOpen, Classified
ReasonThe Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
Root causeComponent design/selection
ActionBetween the dates of July 8, 2026 - July 16, 2026 Fort Defiance Industries (FDI) issued a Urgent Medical Device Recall to affected consignees vi Email. FDI ask consignees to take the following actions: 1. Devices can continue in operation until the fix is implemented. 2. In the interim, users should avoid touching the WRS power connector during operation as a precaution. 3. Please contact FDI and order the upgrade kit complete with instructions to perform this Field Correction. 4. After receiving the kit, please review the instructions and determine if you are competent to perform the upda
DistributionWorldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141009FORT DEFIANCE AUTOMATED STEAM STERILIZER
Fort Defiance Industries, Inc., Dec 16, 2014