Recall Z-2861-2026 · NSU Instrumentation For Clinical Multiplex Test Systems

Nuclein LLC

Brand Name: DASH Rapid PCR Instrument Product Name: DASH Rapid PCR Instrument Model/Catalog Number: SG-0002 Software Version: 6.0.2 Product Description: The DASH Rapid PCR Instrument (DASH Instrument) is an automated, fluorescence-based, reverse-transcription polymerase chain reaction (RT-PCR) instrument that extracts specimens and performs nucleic acid amplified tests (NAAT) in DASH test cartridg

DateAug 3, 2026 (initiated Jun 25, 2025)
StatusOpen, Classified
ReasonDue to a software issue which may result in false co-infection (e.g., Positive for Flu A, when SARS-CoV-2 is present).
Root causeProcess control
ActionOn 06/25/2025, a "MEDICAL DEVICE RECALL" Letter was emailed to customers informing them that, Nuclein identified an incorrect software file on a small number of DASH instruments that creates a possibility for the affected instruments to make incorrect calls that indicate coinfection in the patient tested. On 07/16/2026, the firm emailed an updated "URGENT: MEDICAL DEVICE RECALL" Letter. -Customer were informed regarding updates to the reason for the recall and potential health risks. -Customer were instructed to: All affected DASH Instrument identified in Section 6 (below) have already been
Quantity2
DistributionUS Nationwide distribution in the state of Texas.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K241652DASH® SARS-CoV-2 & Flu A/B Test
Nuclein, LLC, Dec 20, 2024