Recall Z-2860-2026 ยท BSZ Gas-Machine, Anesthesia

Maquet Critical Care AB

Flow Anesthesia Machine Flow-i C20 6677200 Flow-i C30 6677300 Flow-i C40 6677400 Flow-i C20 6888520 Flow-i C30 6888530 Flow-i C40 6888540 Flow-c 6887700 Flow-e 6887900

DateAug 3, 2026 (initiated Jun 24, 2026)
StatusOpen, Classified
ReasonEvents observed where anesthetic agent was not delivered when the vaporizer was activated. Failure to deliver the anesthetic agent may result in the patient receiving insufficient anesthesia and may lead to intraoperative awareness.
Root causeUnder Investigation by firm
ActionOn June 24, 2026, the firm began issuing "Urgent Medical Device Correction" letters via FedEx to customers at the Risk Manager level. Customers were informed that according to the Instructions for Use, this indication signifies that concentration levels are below detectable limits, serving as confirmation that no agent is present: **/gas analyzer has no information to display Customers were given recommended precautions to identify and mitigate the issue, such as confirming that the anesthetic agent is being delivered after activating the vaporizer, and continuously monitoring end-tidal anes
Quantity408 US; 11,342 OUS
DistributionWorldwide - US Nationwide distribution to the states of AL, AR, CA, CO, DC, DE, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, SC, TX, VA, WA, WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191027Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia System
Maquet Critical Care AB, Mar 18, 2020