Recall Z-2911-2026 ยท NIM Stent, Carotid

Contego Medical, Inc

Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40

DateAug 6, 2026 (initiated Jun 29, 2026)
StatusOpen, Classified
ReasonCarotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond issue, may fail to open following sheath retraction; and if this condition were to occur during a procedure, the result may be loss of intended integrated embolic protection.
Root causeProcess control
ActionOn 6/29/2026, recall notices started being delivered to healthcare professionals and risk managers by Medtronic sales representatives who informed of the following: 1) Sales representatives will visit, conduct inventory counts, quarantine, and remove unused affected devices. 2) If replacements are desired, they can be provided. 3) Complete and return the acknowledgment form via email to recall@contegomedical.com Questions can be directed to the firm at recall@contegomedical.com and 919-459-7250 Extension 4.
Quantity5069
DistributionUS Nationwide distribution in the states of IL, FL, NC, WI, KY, AL, VA, DE, NJ, PA, IN, MI, CA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.