ClearanceBrief
Recall Z-2651-2026 ยท JAK System, X-Ray, Tomography, Computed

PHILIPS MEDICAL SYSTEMS

IQon Spectral CT IQon Spectral CT Upgrades

DateJul 27, 2026 (initiated Jun 11, 2026)
StatusOpen, Classified
ReasonThe clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.
Root causeRadiation Control for Health and Safety Act
ActionAn Electronic Product Radiation Correction letter was issued to its customers describing the product, problem and actions to be taken by its customers. To identify the software version of your product: 1. Click the Help button. 2. Select About and the software version is then displayed. The software version begins with v. 3.You may continue to use your system(s) in accordance with the intended use. 4. Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Electronic Product Radiation Correc
Quantity42
DistributionUS Nationwide Distribution including Puerto Rico
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K193454IQon Spectral CT
Philips Medical Systems Nederland B.V., Jan 24, 2020