ClearanceBrief
Recall Z-2906-2026 ยท SFD In Situ Polymerizing Peripheral Nerve Repair Device

TISSIUM SA

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO4010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

DateAug 6, 2026 (initiated Jun 15, 2026)
StatusOpen, Classified
ReasonCurrently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
Root causeIncorrect or no expiration date
ActionOn June 25, 2026, the firm began notifying consignees via direct visit by medical representatives to deliver the Customer Recall Notification Letter in person. Representatives removed affected units from customer shelves, collected completed response forms, and provided conforming replacement product on the same day.
DistributionUS Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.