ClearanceBrief
Recall Z-2895-2026 ยท JIX Calibrator, Multi-Analyte Mixture

Aniara Diagnostica LLC

5-PLASMA Bivalirudin Calibrator, Model Number 5D-46905

DateAug 5, 2026 (initiated Jul 1, 2026)
StatusOpen, Classified
ReasonStability issues with the reconstituted lyophilized human plasma and bivalirudin matrix have been identified. These stability issues lead to a degradation of measurable bivalirudin in the controls, calibrators, and linearity values.
Root causeProcess control
ActionEmail notifications were sent to customers on 6/17/26. Please reply to this email as acknowledgment you have received this initial notification and please include the quantity of each product which is remaining in your possession. A notification letter dated 7/1/26 was sent to customers. Actions to be implemented: " Destroy all stock of 5-PLASMA Bivalirudin Control, 5-PLASMA Bivalirudin Calibrator, & 5-PLASMA AMR Verification Kits - Please permanently deface the box labels and vial labels before disposal. " Complete and return the Awareness Acknowledgement form. " If product has been cons
Quantity7 units
DistributionUS Nationwide distribution in the states of IL, NC, TN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.