Recall Z-2938-2026 ยท GEH Unit, Cryosurgical, Accessories

Erbe USA Inc

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

DateAug 12, 2026 (initiated Jun 26, 2026)
StatusOpen, Classified
ReasonA strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.
Root causeDevice Design
ActionOn June 26, 2026, Erbe USA issued a Urgent: Medical Device Correction update notification to affected consignees via multiple methods ( Email/Website/Hand delivery). ERBE ask consignees to take the following actions: 1. We request the ERBECRYO 2 Unit be made available to the Erbe Representative to install the (V)1.0.4 Software. The Erbe Representative will perform all tasks required to update the software. 2. Confirm acknowledgement with Erbe Representative on-site. 3. Share this notice with all users within your facility. 4.While awaiting software update, there is no change of the Cryosurg
Quantity11226 units
DistributionUS Nationwide distribution including in the states and territories of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michig
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190651ERBECRYO 2 Cryosurgical Unit; ERBECRYO 2 1-pedal footswitch, Extension cable f. ERBECRYO 2 footswitch; Flexible Cryoprob
Erbe Elektromedizin GmbH, Jan 7, 2020