ClearanceBrief
Recall Z-2824-2026 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Instrumentation Laboratory

Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 portable blood analyzer system

DateJul 29, 2026 (initiated May 8, 2026)
StatusOpen, Classified
ReasonPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
Root causeSoftware Design Change
ActionWerfen dba Instrumentation laboratory notified consignees on about 05/08/2026 via letter and/or email. Consignees were instructed to discontinue use of the affected BCR wands immediately, manually enter patient IDs when performing testing on the GEM Premier 5000 and/or 7000 until the BCR wands have been replaced, contact Werfen Service at 1-800-955-9525 to arrange for prompt replacement of the affected BCR wands, post this notification on or near affected GEM Premier 5000 and/or 7000 instruments, and inform all laboratory staff of this notification. Consignees were also instructed to forward
Quantity67
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K203790GEM Premier 5000
Instrumentation Laboratory CO, Aug 20, 2021