ClearanceBrief
Recall Z-3052-2026 ยท FDE Laparoscopy Kit

Medline Industries, LP

Medline Convenience Kits: 1) COLORECTAL ME, Model Number: DYNJ906437D

DateAug 31, 2026 (initiated Jun 12, 2026)
StatusOpen, Classified
ReasonThe kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patient
Root causeNonconforming Material/Component
ActionAn URGENT MEDICAL DEVICE RECALL notification letter dated 6/12/25 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-097-FGX1
Quantity176 kits
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.