ClearanceBrief
Recall Z-2744-2026 ยท FTC Light, Ultraviolet, Dermatological

Daavlin Distributing Company

Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders

DateJul 15, 2026 (initiated Mar 5, 2026)
StatusOpen, Classified
ReasonCalibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Root causeSoftware Design Change
ActionDaavlin Distributing Company initially notified its consignees about the issue beginning on 3/5/2026 via telephone following with by an email. The firm a follow up URGENT MEDICAL DEVICE Advisory notice on 4/9/2026 with the same information via US mail and email. On 7/10/2026 the firm sent an URGENT MEDICAL DEVICE RECALL notice to all non-responding consignees. The notice explained the issue, potential risk to health, and requested the following: "" Verify your device is operating ClearLink Controller Software Version 3.17 as follows: a) Ensure your machine is plugged in and on b) Locate
Quantity212 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K182215ClearLink Controlled Phototherapy Equipment
Daavlin Distributing Company, Oct 29, 2018