Recall Z-2856-2026 ยท KNN Components, Wheelchair

MUNEVO GMBH

munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System

DateJul 31, 2026 (initiated Jun 19, 2026)
StatusOpen, Classified
ReasonA software issue may cause the wheelchair's driving mechanism to fall out of calibration. When this occurs, the device may experience unexpected steering drift or imprecise control while driving the wheelchair. The wheelchair may veer from its intended path, which may cause in low-speed collision, tipping, or injuries to user or bystanders.
Root causeSoftware Design Change
ActionOn June 19, 2026, the firm issued an "Urgent Medical Device Recall" letter to consignees, which detailed the hazard, mandatory boot-up safety protocols, and manual override steps. The firm has developed software patch Version 12.1.7 (v1), and 12.2.6 (R-net) (released patch on June 17 2026) which corrects the profile retention database loop, ensuring the device consistently locks onto the last saved user profile during boot. This software patch was released to all users on June 17, 2026. Customers should update the software as soon as possible. Until the update is installed, do not activate n
Quantity19
DistributionWorldwide - US Nationwide distribution in the states of California, Florida, North Carolina, Texas, Utah, and Washington. The countries of United Kingdom, Spain, Italy, Germany, and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.